- Quantification of Intracranial Aneurysm Volume Pulsation with 7T MRI
Tenunruptured aneurysms in 9 patients were studied using a high-resolution 3D gradient-echo sequence with cardiac gating. Semiautomatic segmentation was used to measure aneurysm volume per cardiac phase. Aneurysm pulsation was defined as the relative increase in volume between the phase with the smallest volume and the phase with the largest volume. The accuracy and precision of the measured volume pulsations were addressed by digital phantom simulations and a repeat image analysis. In Stage II, the imaging protocol was optimized and 9 patients with 9 aneurysms were studied with and without administration of a contrast agent. Mean aneurysm pulsation in Stage I was 8%, with a mean volume change of 15 mm3. The artifactual volume pulsations measured with the digital phantom simulations were of the same magnitude as the volume pulsations observed in the patient data. Volume pulsation quantification with the current imaging protocol on 7T MR imaging is not accurate due to multiple imaging artifacts.