- Toward Better Understanding of Flow Diversion in Bifurcation Aneurysms
Fundamental electric and fluid dynamics principles were applied to generate equations describing the relationships between changes in flow and the degree of vessel coverage in settings of variable collateral support to the jailed territory. The authors studied a simplified hypothetic system with minimum assumptions to generate the most conservative outcomes. Up to 30% metal coverage of any branch territory is very likely to be well-tolerated regardless of device or artery size or the availability of immediate collateral support, provided that no acute thrombus forms to further reduce jailed territory perfusion.