Table 1:

DWI sequence parameters

SequencetseDWIrsDWI
TR (ms)20004000
TE (ms)10366 (and 91 for phase-correction scan)
Voxel size (mm³)1.1 × 1.5 × 31.4 × 1.4 × 3
FOV (mm²)220230
FOV in-phase direction100%65%
Phase directionAnterior-posterior (axial), right to center (coronal)Right to center (axial and coronal)
Phase resolution75%100%
Partial Fourier50% (phase)87.5% (readout)
Matrix192 × 144160 × 104
Section distance10%10%
No. of slices13 (axial)11 (coronal)13 (axial)11 (coronal)
Parallel imagingGRAPPA ×2GRAPPA ×2
Bandwidth (Hz/pixel)554977
Echo spacing (ms)4.480.36
Readout segments15
Flip angle150°180°
b-values (s/mm2)10000, 1000
Averages101 (for b=0 s/mm²), 2 (for b=1000 s/mm²)
Diffusion mode3D diagonal4-scan trace
Diffusion schemeBipolarBipolar
Acquisition time (minute:second)4:22 (axial), 3:42 (coronal)3:06
  • Note:—GRAPPA indicates generalized autocalibrating partially parallel acquisition.