Table 4:

Prediction of LVO by the Viz LVO system

System LVO DetectionSensitivity95% CISpecificity95% CINPV95% CIPPV95% CIAccuracy95% CI
Entire cohort (n = 1167)0.810.74–0.910.960.95–0.970.990.98–0.990.650.55–0.740.940.92–0.96
Stroke protocol subgroup (n = 404)0.820.71–0.890.900.86–0.930.960.93–0.980.640.53–0.730.890.86–0.94
  • Note:—NPV indicates negative predictive value.