Table 3:

Comparative performance of drill systems in sclerotic lesions based on densitya

Density (HU)Diagnostic YieldCrush ArtifactsRadiation Dose (mGy × cm2)Scanning Time (min)
Manual
 250–70088.5 (23/26)3.8 (1/26)104445.1
 >70085.7 (12/14)2.1 (3/14)131746.1
P value.583.115.3.862
Battery-powered
 250–70089.3 (25/28)0 (0/28)88934.6
 >70085 (17/20)20 (4/20)75241
P value.562.025b.38.026b
  • a Percentage followed by number of cases/total number of cases in parentheses for diagnostic yield and crush artifacts. Mean values are listed for radiation dose and scanning time.

  • b Significant P value (P < .05).