Table 1:

Acute stroke MRI acquisition parameters

ParameterAcute Stroke MRI
SS EPI-DWISWIDCE-MRA (TWIST)DSC-MRPT2-FLAIR
TR/TE (ms)4800/7128/202.62/0.951600/309000/95
TI (ms)2500
Turbo factor21
Flip angle15°21°90°150°
Bandwidth (Hz/pixel)16721207801446206
Slice thickness/gap (mm)5/22/01.2/05/25/2
FOV (mm)240 × 240240 × 195400 × 300240 × 240240 × 217
Matrix size (mm)130 × 130384 × 156320 × 182128 × 128256 × 174
B-value (s/mm2)0, 1000
Measurements1130601
GRAPPA223 × 22
Temporal resolution (sec)1.61.6
Dynamic reconstruction modeForward share
K-space sampling (center/periphery) (%)15/20
  • Note:—SS indicates single-shot; TWIST, time-resolved imaging with stochastic trajectories; GRAPPA, generalized autocalibrating partially parallel acquisition.