Table 4:

2D contrast protocol

NameTOF MRA LocalizerOblique T1WI LocalizerPrecontrast and Postcontrasta T1WI3D-TOFMPRAGE
Plaque featureLocalize arteryLocalize bifurcationPlaque burden and distribution, LRNCStenosis, ulceration, calcificationIntraplaque hemorrhage
SequencebFFE/SPGRTSE/FSETSE/FSEFFE/SPGRIR-TFE/IR-FSPGR
Image mode2D2D2D3D3D
Scan planeAxialObliqueAxialAxialAxial
TR (ms)2520008002413
TE (ms)4810MinMin
FOV (cm)16 × 1216 × 1616 × 1616 × 1616 × 16
Resolution (mm2)0.63 × 0.830.63 × 0.630.63 × 0.630.63 × 0.630.63 × 0.63
Slice thickness (mm)2222/−12/−1
No. of slices326164040
Blood suppressionSaturation–veinsMSDEQIRcSaturation–veinsNod
Special parametersFlip angle = 40°Echo-train = 14; 3 slices/TREcho-train = 10Flip angle = 20°Flip angle = 15°, turbo factor = 30, TI = 500 ms, IRTR = 800 ms
Fat suppressionNoYesYesNoWater excitation
  • Note:—QIR indicates quadruple inversion recovery.

  • a Clinical CE-MRA per institutional protocol can be used before postcontrast sequence.

  • b Siemens/Philips/GE acronyms.

  • c Double inversion recovery can be used, but blood suppression may be incomplete, and inversion time is a variable dependent on patient and contrast bolus.

  • d Pulse gating not required for any sequence.