Table 1:

Radiation dose indices, noise, and CNR of temporal bone axial scans with different detector collimations

Detector Collimation (mm)CTDIvol(mGy)DLP (mGy · cm)Noise (HU)CNR
Brain StemSphenoid
2 × 0.62547.21898.677.941.416.1
12 × 0.62520.8935.618.715.060.5
16 × 0.62522.6902.882.943.415.1
40 × 0.62524.21211.575.334.617.2
64 × 0.62519.4775.680.449.614.8
  • Note:—When the 12 × 0.625-mm collimation was selected, the system could not choose 180 mAs/section. The closest tube mAs, 165 mAs/section, was selected. In addition, the 12 × 0.625-mm collimation used the system's “detailed resolution” setting and got substantially different noise and CNR, with no comparability with other collimations. The 12 × 0.625-mm detector collimation was not available in helical mode and was not included in the comparison.