Table 3:

Inter- and intraobserver agreement for a WBCE-3D TSE T1WI and OCE-2D coronal T1WI when detecting optic nerve or optic chiasm enhancement

InterobserverIntraobserver
OCE-2D T1WI κ (95% CI)WBCE-3D T1WI κ (95% CI)OCE-2D T1WI κ (95% CI)WBCE-3D T1WI κ (95% CI)
Optic nerve enhancement0.80 (0.75–0.85)0.65 (0.60–0.71)0.75 (0.70–0.80)1 (1–1)
Localization of the enhancement
 Bilateral0.90 (0.78–1)0.53 (0.34–0.71)0.75 (0.58–0.92)1 (1–1)
 Intraorbital segment0.76 (0.70–0.82)0.63 (0.57–0.70)0.81 (0.76–0.87)0.81 (0.76–0.87)
 Canalicular segment0.64 (0.56–0.73)0.72 (0.65–0.79)0.71 (0.64–0.79)0.68 (0.58–0.75)
 Cisternal segment0.58 (0.45–0.72)0.54 (0.37–0.72)0.49 (0.34–0.65)0.81 (0.76–0.87)
 Optic chiasm0.44 (0.15–0.73)0.73 (0.51–0.96)0.33 (0.04–0.61)1 (1–1)