Table 3:

Reproducibility of volume estimation of WM, GM, CSF, BPV, and ICV using NeuroQuant and SyMRI, between 1.5T and 3Ta

T1-Weighted NQ 1.5T-3TsyT1WI NQ 1.5T-3TSyMRI 1.5T-3T
MeanBiasSDrMeanBiasSDrMeanBiasSDr
WM66127b650.8361358b230.975470390.89
GM622−6580.8360955b140.98669−7330.95
CSF368–41b340.76359–45b220.882361160.97
BPV1284211190.841223113b310.98125821b131.00
ICV1652−201200.86158168b200.99149421b91.00
  • a Expressed as mean, bias, and SD of the difference and correlation coefficient r.

  • b Significant (P < .05).