PT - JOURNAL ARTICLE AU - Pearly Ti, J. AU - Yeo, L. AU - Anil, G. TI - Can a Stent Retriever Damage the JET 7 Reperfusion Catheter? AID - 10.3174/ajnr.A6804 DP - 2020 Dec 01 TA - American Journal of Neuroradiology PG - 2317--2319 VI - 41 IP - 12 4099 - http://www.ajnr.org/content/41/12/2317.short 4100 - http://www.ajnr.org/content/41/12/2317.full SO - Am. J. Neuroradiol.2020 Dec 01; 41 AB - SUMMARY: We encountered 2 separate instances of damage to JET 7 reperfusion catheters when they were used in conjunction with a stent retriever during mechanical thrombectomy. On both occasions, after 1 or 2 passes with a stent retriever, we found that the distal end of the catheter was frayed and it ballooned up on flushing with saline. This mechanical failure could potentially lead to serious complications; hence, it should be shared with fellow neurointerventionalists.DACdistal access catheters